25G Sampling Oscilloscope
Classification:
DCA25021 optical sampling oscilloscope is used for testing of transmitter performance indicators. It supports 4-channel optical signal input, NRZ and PAM4 signal testing. The optical channel bandwidth is 30GHz with a variety of typical filters. The electrical channel bandwidth is up to 35GHz.
- Application
- Characteristic
- Specifications
- Download
-
Application
1. Optical module eye diagram parameter test
2. Optical transmitter performance testFeatures
1. High resolution, high sampling rate ADC
2. Electrical channel bandwidth up to 35GHz
3. Optical channel supports filter options of different standard rates
4. Optical receiver supports single-mode and multi-mode interfaces
5. Support NRZ and PAM4 eye diagram testing
6. Rapid NRZ testing of eye diagram full parameters in only 1s
7. SSPRQ pattern locking in only 6sTest
1. Support 4-channel 25.78Gbps NRZ optical signal eye diagram test at the same time
2. Support 4-channel 26.5625GBaud PAM4 optical signal eye diagram test at the same time

Specification

-
• High-resolution, high-sampling-rate ADC
• Electrical channel bandwidth up to 50 GHz
• Optical channel supports different rate filter options
• Optical receiver supports single-mode and multi-mode interfaces
• Supports NRZ and PAM4 eye diagram testing
• Fast eye diagram testing with all parameters in just 0.55 seconds
• Software running interface is as follows:
-
Optical Channel Parameters
Reference Value
Optical Channel Bandwidth
30GHz
Wavelength Range
850nm~1650nm, Standard Wavelengths: 850/1310/1550nm
Sensitivity
-12dBm
Average Power (Consistency)
±0.1dB
Extinction Ratio (Consistency)
±0.1dB
Template Margin (Consistency)
±2.5%
Maximum Input (Non-destructive Peak Power)
+5dBm
Maximum Input (Linear Test Range)
0dBm
Monitoring Average Power Measurement Range
-17dBm~+3dBm
Input Return Loss
30dB
Fiber Optic Access
62.5/125um FC/UPC
Electrical Channel Parameters
Reference Value
Maximum Input Bandwidth
50GHz
Maximum Input Range
1Vpp
RMS Noise
<1mV
Sensitivity
<50mV
Input Impedance
50Ω
Return Loss
<-18dB
Input Interface
3.5mm/2.92mm(Female)
Trigger Parameters
Reference Value
Clock Input Bandwidth
100MHz~24GHz
Clock Input Sensitivity
<50mV
Trigger Maximum Input Signal
≤±2V
Trigger Input Impedance
50Ω
Trigger Input Interface
3.5mm/2.92mm(Female)
Time Base Interval Accuracy
<50fs
Sampling Parameters
Reference Value
Number of Sample Points
2048 points/Waveforms
Cumulative Waveform Range
1~100000 Waveforms (More waveform ranges can be customized)
Refresh Rate
200Waveforms/S
Overall Machine Parameters
Reference Value
Number of Channels
Optical Channel x1; Electrical Channel x1
Filter
NRZ: 1.229/1.244/1.25/2.5/10Gb/s,2.458/2.488/2.5Gb/s,3.072/3.125Gb/s,6.144/6.25Gb/s
PAM4: 26.5625 Gbaud、28.9 Gbaud
25.78 Gb/s(25/50/100 Gb Ethernet), 26.56 Gb/s(400 Gb Ethernet), 27.95 Gb/s(OTU4)
28.05 Gb/s(32× Fiber Channel)
More rate filters can be customized
Temperature Monitoring
Supported
Key words:
Focus on us
Tel
086-27-8990 8766